Slice sensitivity profile and image pixel noise of multi-slice spiral CT in comparison with single slice spiral CT


     Related Videos
Class II Slice Technique
Sudhir Venkatesh - A Slice of Gang Life
How to slice and export a website design from Photoshop
Slice and Export a Website Layout HTML Photoshop Tutorial!

     Related Hubpages

    •  Doc. Url:    Embed Code: 

    • PubMed  status
      (0) (0 Votes)
      Views: (1)   Date: (1999-10-16)   Pages: ()
    • Author:  E Grabbe  

    • Abstract:  PURPOSE: Presentation and evaluation of slice sensitivity profile and pixel noise of multi-slice CT in comparison to single-slice CT. METHODS: Slice sensitivity profiles and pixel noise of a multi-slice CT equipped with a 2D matrix detector array and of a single-slice CT were evaluated in phantom studies. RESULTS: For the single-slice CT the width of the slice sensitivity profiles increased with increasing pitch. In spite of a much higher table speed the slice sensitivity profiles of multi-slice CT were narrower and did not increase with higher pitch. Noise in single-slice CT was independent of pitch. For multi-slice CT noise increased with higher pitch and for the higher pitch decreased slightly with higher detector row collimation. CONCLUSIONS: Multi-slice CT provides superior z-resolution and higher volume coverage speed. These qualities fulfill one of the prerequisites for improvement of 3D postprocessing.