Abstract: A new quantitative antiglobulin consumption (QAC) test for the measurement of platelet-associated IgG is described. In this test washed platelets are incubated with anti-IgG at a final dilution of 1:2 million. The unneutralized fraction of anti-IgG remaining in solution is then measured with an Autoanalyzer and soluble IgG is used for calibration. The dose-response curves depicting the percent neutralization of anti-IgG by platelets and by soluble IgG were compared in detail and found to be nearly identical, indicating that platelet-associated IgG can be accurately quantitated by this method. The mean IgG values were 2287 molecules/platelet for normal adults and 38,112 molecules/platelet for ITP patients. The Autoanalyzer QAC test is a sensitive and reproducible assay for the quantitation of platelet-associated IgG.